Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
6182
Unsigned / 2
Backup
Command to overwrite
the existing backup da‐
ta sets.
This ensures that the
current sensor serial
number and transmitter
serial numbers are up‐
dated in the backup da‐
taset. This enables the
automatic backup.
-
● 1: Overwrite existing
backup files
Write on‐
ly
6099
Unsigned / 2
Restore
Command to restore
backup data. Coding is
bitwise.
-
● Bit 0: Restore sensor
setup parameters
● Bit 1: Restore sensor
application setup
parameters
● Bit 2: Restore
transmitter
application setup
parameters
● Bit 3: Restore
customer transmitter
configuration setup
parameters
● Bit 4: Restore
totalizers
● Bit 5: Restore system
setup
● Bit 6: Restore
customer sensor
configuration setup
parameters
Write on‐
ly
4.10.10.2
Data logging
Table 4-93
Data logging
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
11400
Unsigned / 2
Enable / disable Controls data logging.
0
● 0: Disabled
● 1: Enabled
Read /
write
11401
Unsigned / 2
Data logging
mode
Controls data logging
mode.
0
● 0: Log instantaneous
● 1: Log average value
Read /
write
11402
Unsigned / 4
Logging interval Defines interval for log‐
ging. Granularity is 10
ms.
500
[x 10 ms]
1 - 8640000
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
176
Function Manual, 06/2018, A5E39931813-AB