GPIB language is set as SCPI by default and cannot be changed for the
R&S
FPS.
Notes and Conditions
In connection with the GPIB interface, note the following:
●
Up to 15 instruments can be connected
●
The total cable length is restricted to a maximum of 15 m or 2 m times the
number of devices, whichever is less; the cable length between two instru-
ments should not exceed 2 m.
●
A wired "OR"-connection is used if several instruments are connected in par-
allel.
●
Any connected IEC-bus cables should be terminated by an instrument or con-
troller.
5.1.3
USB Interface
For remote control via the USB connection, the PC and the instrument must be
connected via the USB type B interface. A USB connection requires the VISA
library to be installed. VISA detects and configures the R&S instrument automati-
cally when the USB connection is established. You do not have to enter an
address string or install a separate driver.
USB address
The used USB address string is:
USB::<vendor ID>::<product ID>::<serial number>[::INSTR]
where:
●
<vendor ID> is the vendor ID for Rohde & Schwarz (0x0AAD)
●
<product ID> is the product ID for the Rohde & Schwarz instrument
●
<serial number> is the individual serial number on the rear of the instrument
Table 5-2: Product IDs for R&S
FPS
Instrument model
Product ID
FPS4
F9
FPS7
FA
FPS13
FB
Remote Control Interfaces and Protocols