Preparing for Use
R&S
®
FPS
26
Getting Started 1319.3362.02 ─ 07
data created during operation. If necessary, the solid-state drive can be removed
from the R&S
FPS and locked in a secure place to protect any classified data it
may contain.
Redirecting storage to volatile memory
Alternatively, to avoid storing any sensitive data on the R&S
FPS permanently,
the
secure user mode
was introduced (option R&S
FPS-K33). In secure user
mode the instrument’s solid-state drive is write-protected so that no information
can be written to memory permanently. Data that the R&S
FPS normally stores
on the solid-state drive is redirected to volatile memory instead, which remains
available only until the instrument is switched off. This data includes:
●
Windows 7 operating system files
●
Firmware shutdown files containing information on last instrument state
●
Self-alignment data
●
General instrument settings such as the IP address
●
Measurement settings
●
User data created during operation
●
Any data created by other applications installed on the R&S
FPS, for example
text editors (Notepad), the Clipboard, drawing tools, etc.
Users can access data that is stored in volatile memory just as in normal opera-
tion. However, when the instrument’s power is switched off, all data in this mem-
ory is cleared. Thus, in secure user mode, the instrument always starts in a
defined, fixed state when switched on.
To store data such as measurement results permanently, it must be stored to an
external storage device, such as a memory stick.
Limited storage space
The volatile memory used to store data in secure user mode is restricted to
256
MB. Thus, a "Memory full" error can occur although the hard disk indi-
cates that storage space is still available.
Storing required data permanently
Any data that is to be available for subsequent sessions with the R&S
FPS must
be stored on the instrument permanently,
before activating the secure user mode
.
Protecting Data Using the Secure User Mode
Summary of Contents for R&S FPS
Page 25: ......