Operating the Instrument in Manual Mode
R&S
®
FPS
104
Getting Started 1319.3362.02 ─ 07
Table 6-9: Status bar information - color coding
Color
Type
Description
red
Fatal
A serious error occurred in the application; regular operation is no
longer possible.
red
Error
An error occurred during a measurement, e.g. due to missing data
or wrong settings, so that the measurement cannot be completed
correctly.
orange
Warning
An irregular situation occurred during measurement, e.g. the set-
tings no longer match the displayed results, or the connection to an
external device was interrupted temporarily.
gray
Information
Information on the status of individual processing steps.
gray
Message
An event or state has occurred that may lead to an error during fur-
ther operation.
green
No errors
No messages displayed.
If any error information is available for a measurement channel, the icon
is displayed next to the channel name.
This is particularly useful when the MultiView tab is displayed, as the status
bar in the MultiView tab always displays the information for the currently
selected measurement only.
The following keywords are used:
IF OVLD
Overload of the IF signal path after the input mixer.
●
Increase the reference level.
INPUT OVLD
The signal level at the RF input connector exceeds the maximum.
The RF input is disconnected from the input mixer to protect the device. In
order to re-enable measurement, decrease the level at the RF input con-
nector and reconnect the RF input to the mixer input.
LOUNL
Error in the instrument's frequency processing hardware was detected.
NO REF
Instrument was set to an external reference but no signal was detected on
the reference input.
OVEN
OCXO reference frequency (option R&S
FPS-B4) has not yet reached its
operating temperature. The message usually disappears a few minutes
after power has been switched on.
RF OVLD
Overload of the input mixer.
●
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)