Instrument Tour
R&S
®
FPS
34
Getting Started 1319.3362.02 ─ 07
Chapter 5, "Controlling the R&S
3.2.5
TRIGGER INPUT / OUTPUT
The female TRG IN connector for external trigger or gate input is used to control
the measurement by means of an external signal. The voltage levels can range
from 0.5 to 3.5 V. The default value is 1.4 V. The typical input impedance is 10
kΩ.
The female BNC TRG AUX connector can be used to receive a second external
signal or to provide a signal to another device. The signal is TTL compatible (0 V /
5 V). You can control the connector usage in the "Trigger" settings (TRIG key).
3.2.6
IF/VIDEO OUTPUT
The female BNC connector can be used for various outputs:
●
Intermediate frequency (IF) output of approximately 20 MHz
●
Video output (1V)
Which output is provided is defined in the software (INPUT/OUTPUT key). For
details see the User Manual.
3.2.7
DISPLAY PORT and DVI
You can connect an external monitor or other display device to the R&S
FPS to
operate the instrument from a remote PC (see also
Two different types of connectors are provided for this purpose:
●
DISPLAY PORT
●
DVI (Digital visual interface)
3.2.8
LAN
The LAN interface can be used to connect the R&S
FPS to a local network for
remote control, printouts and data transfer. The assignment of the RJ-45 connec-
Rear Panel View