Operating the Instrument in Manual Mode
R&S
®
FPS
101
Getting Started 1319.3362.02 ─ 07
Rm
RMS detector
Trace Mode
Sweep mode:
Clrw
CLEAR/WRITE
Max
MAX HOLD
Min
MIN HOLD
Avg
AVERAGE (Lin/Log/Pwr)
View
VIEW
Norm/NCor
Correction data is not used.
6.2.4
Marker Information
Marker information is provided either in the diagram grid or in a separate marker
table, depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x and y-axis positions of the last 2 markers or delta
markers that were set, if available, as well as their index are displayed. The value
in the square brackets after the index indicates the trace to which the marker is
assigned. (Example: M2[1] defines marker 2 on trace 1.) For more than 2 mark-
ers, a separate marker table is displayed beneath the diagram by default.
Marker information in marker table
In addition to the marker information displayed within the diagram grid, a separate
marker table may be displayed beneath the diagram. This table provides the fol-
lowing information for all active markers:
Type
Marker type: N (normal), D (delta), T (temporary, internal)
Ref
Reference (for delta markers)
Trc
Trace to which the marker is assigned
X-value
x-value of the marker
Y-value
y-value of the marker
Func
Activated marker or measurement function
Func. Result
Result of the active marker or measurement function
Understanding the Display Information