Operating the Instrument in Manual Mode
R&S
®
FPS
99
Getting Started 1319.3362.02 ─ 07
sured values. This information is displayed in gray font and only when applicable
for the current measurement, as opposed to the channel-specific settings that are
always displayed.
The following types of information may be displayed, if applicable.
Table 6-8: Common settings displayed in the channel bar
SGL
The sweep is set to single sweep mode.
Sweep Count
The current signal count for measurement tasks that involve a specific number
of subsequent sweeps
(see "Sweep Count" setting in "Sweep settings" in the User Manual)
TRG
Trigger source
(for details see "Trigger settings" in the User Manual)
●
EXT
: External
●
IFP
: IF power (+trigger bandwidth)
●
RFP
: RF power
●
VID
: Video
6dB/RRC/
CHN
Filter type for sweep bandwidth
(see "Bandwidth settings" in the User Manual)
YIG Bypass
The YIG filter is deactivated.
PA
The preamplifier is activated.
GAT
The frequency sweep is controlled via the TRIGGER INPUT connector.
TDF
A transducer factor is activated.
75
Ω
The input impedance of the instrument is set to 75
Ω
.
FRQ
A frequency offset
≠
0 Hz is set.
DC/AC
An external DC or AC calibration signal is in use.
Understanding the Display Information