Graphic Measurement Modes
R&S
®
EVSF1000
95
User Manual 1178.6410.02 ─ 05
●
: the x-axis starts with the start frequency and ends with the stop fre-
quency
Center
Configures the center frequency. The configured value is displayed as "Center" in the
measurement settings area.
Remote command:
Span
Configures the frequency span, which must be at least 100
kHz. The configured value
is displayed as "Span" in the measurement settings area.
Remote command:
Start
Configures the start frequency of the x-axis. The configured value is displayed as
"Start" in the measurement settings area.
Remote command:
Stop
Configures the stop frequency of the x-axis. The configured value is displayed as
"Stop" in the measurement settings area.
Remote command:
7.1.2
Configuring the Bandwidth
Access VNC
: [e] > "F1"
Resolution Bandwidth (RBW Mode/Res BW)
............................................................... 95
Resolution Bandwidth (RBW Mode/Res BW)
Defines the resolution bandwidth (RBW) of the spectrum. The resolution bandwidth
defines the minimum frequency separation at which the individual components of a
spectrum can be distinguished. Small values result in high precision, as the distance
between two distinguishable frequencies is small. Higher values decrease the preci-
sion, but increase measurement speed.
The RBW can be determined automatically by the R&S
EVSF1000, or manually.
AUTO mode
The RBW is determined automatically according to the sweep time
and frequency range.
RF Spectrum Mode (Option R&S
EVSG-K10)