
Common Instrument Settings
R&S
®
EVSF1000
172
User Manual 1178.6410.02 ─ 05
Time
Sets the time of the internal clock in the format
hh.mm.ss
.
Example:
12.08.23
Boot on Power Up
Automatically switches on the R&S
EVSF1000 when it is connected to an external
power supply (if not in operation already). This setting can not be changed on the
R&S
EVSF1000.
Remote command:
Flash size / Flash used
Indicates the size of the flash memory in the R&S
EVSF1000, and the percentage cur-
rently in use.
11.5
Instrument Configuration and Operating Status Set-
tings
The following settings concern the instrument operating status, and hardware and soft-
ware configuration.
............................................................................. 174
Auto-Calibration (Cal)
Access VNC
: [n] > [F8] > [F1]
Auto-calibration corrects changes in the attenuation characteristics of the internal RF
signal paths (on both receiver boards, if available) using an internal calibration signal.
These changes can occur due to the aging process or temperature drifts. The calibra-
tion optimizes the accuracy of the level measurements.
The "Setup - CAL" display shows the status for each calibration step for the most
recent calibration. The first line of the result table indicates the date and time of the
auto-calibration, and the temperature during the auto-calibration.
Note:
Before starting the auto-calibration, the instrument should be in operation for at
least 15 minutes to ensure the instrument has reached its operating temperature (for
details, refer to the data sheet).
During the auto-calibration procedure, do not apply any input signal to the
R&S
EVSF1000.
Instrument Configuration and Operating Status Settings