SETUP:TIMESYNC?
SETUP:TIMESYNC
<Source>
Defines or queries the time source with which the R&S
EVSF1000 clock is synchron-
ized. If the R&S
EVSF1000 clock differs from the selected synchronization source time
by more than 1000
ms, the R&S
EVSF1000 clock is adjusted accordingly.
Parameters:
<Source>
OFF
No synchronization is performed for the internal R&S
EVSF1000
clock.
GPS
Synchronization with a connected GPS (see the main
R&S
EVSF1000 User Manual).
Example:
SETUP:TIMESYNC OFF
12.1.7
System Settings
The following commands are required to configure basic system settings.
................................................................................................ 199
..........................................................................................200
FACTORY_PRESET
This command restores the factory default settings, the IP-settings and Data logger
lists will be not modified.
Return values:
<State>
READY.
The command was executed successfully.
Manual operation:
See
SETUP:ENERGYSAVER
<Value>
This command specifies the energy saver properties.
Parameters:
<Value>
ON
The energy saver is on.
OFF
The energy saver is off.
1 to 60
The energy saver is started after the defined time.
Default unit: min