Remote Commands
R&S
®
EVSF1000
228
User Manual 1178.6410.02 ─ 05
AM5
This command queries the AM modulation of 1300 Hz signal.
Return values:
<%>
Provides the AM5 modulation in %.
Manual operation:
See
AM6
This command queries the AM modulation of 400 Hz signal.
Return values:
<%>
Provides the AM6 modulation in %.
Manual operation:
See
AM7
This command queries the AM modulation of the identifier.
Return values:
<%>
Provides the AM7 modulation in %.
12.3.3
Retrieving ILS MB Identifier Results
The following commands are required to retrieve identifier results from ILS MB mea-
surements.
ID_F400_DASHLEN
Queries the dash time of 400 Hz signal in "ILS" MB mode.
Return values:
<Time>
Length of time a dash for the outer marker.
Default unit: ms
Manual operation:
See
ID_F1300_DASHLEN
Queries the dash time of 1300 Hz signal in "ILS" MB mode.
ILS Marker Beacon (MB) Mode