Common Instrument Settings
R&S
®
EVSF1000
173
User Manual 1178.6410.02 ─ 05
To start the auto-calibration, select "Start Autocal " and press the [Enter] key.
The auto-calibration procedure is ready after a few minutes and in the result table, all
entries must have the status "OK".
Figure 11-2: Successful auto-calibration
If an error occurs during the auto-calibration process, switch the instrument off and on
again and repeat the auto-calibration process. If the error still exists, contact the
Rohde
&
Schwarz service.
It is recommended that you perfom an auto-calibration every 2 months or if the differ-
ence of the environment temperature changes by more than 10
°C.
Remote command:
Error Log
Access VNC
: [n] > [F8] > [F2]
The last 100 instrument status messages or errors displayed in the status bar during
operation of the R&S
EVSF1000 are also stored in an error log file on the instrument.
Thus, if problems occur, you can check the error log for irregular behavior or failures.
Instrument Configuration and Operating Status Settings