Numeric Measurement Modes
R&S
®
EVSF1000
85
User Manual 1178.6410.02 ─ 05
Figure 6-19: RF level display
The following results are provided:
●
Numeric power level in dBm ("Lev")
For measurements on two frequencies: individual and sum power levels
●
Measured frequency offset to the nominal frequency in kHz
●
Numeric frequency offset of measured power
●
Bargraph indicating the power and a color-coded overload state, where:
–
red
: overload state, check message
–
yellow
: power approaching overload state
–
green
: power in a valid range
●
If applicable: overload messages
Overload messages
The following messages indicate an overload:
●
"RF Overload"
Overload of the input mixer or of the analog IF path.
●
"IF Overload"
Overload of the IF signal.
●
"ADC Overload"
The dynamic range of the AD-converter is exceeded (clipping).
A combination of these overloads is also possible.
In all cases, set the RF attenuation to normal or low distortion (for RF input), or reduce
the input level.
Remote commands to retrieve results:
COM Mode (Option R&S
EVSG-K6)