Numeric Measurement Modes
R&S
®
EVSF1000
83
User Manual 1178.6410.02 ─ 05
"Norm"
0 dB
Provides a normal sensitivity.
"Low Dist"
15 dB attenuation
Provides a low sensitivity. Suitable when analyzing a nearby signal, to
avoid overload due to high-level signals.
Remote command:
Transducer Correction
Configures the level correction. The specified value is added to the measured power
levels to compensate for an inherent offset by the measurement setup, for example the
antenna.
6.3.3.4
Bandwidth (BW)
Access VNC
: [e] > [F1]
The bandwidth determines the frequency range for which the measurement is per-
formed on the IF signal.
6.3.3.5
Setting the Measurement Time (MTime)
Access VNC
: [d]
The measurement time determines the interval at which new measurement results are
displayed. Internally, values are captured every 100
ms, that is: 10 per second. (With
the High Measurement Rate option R&S
EVSG-K22 installed, the internal capture rate
is 10
ms, that is: 100 per second.) If the defined measurement time is longer, the val-
ues captured internally in that interval are averaged and only the average value is dis-
played.
Note that the measurement time determines the minimum interval for values to be
stored during recording (see
Enter the time in milliseconds.
Remote command:
6.4
COM Mode (Option R&S
EVSG-K6)
Access VNC
: [m] > Down arrow key
COM Mode (Option R&S
EVSG-K6)