Setting the Output Conditions
ZW User's Manual
87
4
Settings for
Func
tion
3
Select [Actual scaling (two points)], and
click [Next >].
The [Scaling] popup menu display changes to
[1st point settings].
4
Set correction of the 1st point.
Click [STOP] to fix the [Current value], set
[Set value] and click [Next >].
The [Scaling] popup menu display changes to
[2nd point settings].
5
In the same way, set correction of the 2nd
point.
The [Scaling] popup menu display changes to
[Direction settings]. When the change direction
is set and [Next>] is clicked, the [Scaling] popup
menu display changes to [Confirm the scaling
parameters], and the span and offset values are
displayed.
6
Click [OK].
2-point scaling can also be executed by the operating keys on the Sensor Controller.
Performing 2-point Scaling p.242
The current measured value is displayed at
[Current value], and is refreshed at 100 ms
cycles. When [STOP] is clicked, refreshing of the
[Current value] display stops. Clicking the
[STOP] button again resumes display refreshing.
Note
Note