No-protocol Connection
178
ZW User's Manual
Command List
This table lists no-protocol communications commands.
The available commands are listed as follows.
Command
name
Format
Return value
*1
Description
Pages
MS
MS <Task number>
<Delimiter>
<Measured value>
<Delimiter>
Acquires the current measured value.
If the <task number> is omitted, the
measured value displayed is acquired.
If "4" is set for the <task number>, the
measured values for all tasks are obtained.
JG
JG <Task number>
<Delimiter>
<Judgment result>
<Delimiter>
Acquires the judgment result of the specified
task.
If the <task number> is omitted, the result of
the task of which the result is currently
displayed is acquired.
If "4" is set for the <task number>, the
judgment results for all tasks are obtained.
DG
DG <unit number>
<Data number>
<Delimiter>
<Data>
<Delimiter>
Acquires the measurement data and setting
data of the processing unit.
DS
DS <unit number>
<Data number>
<Measured value> <Delimiter>
OK <Delimiter>
Change the setting data of the processing
unit.
BG
BG <Delimiter>
<Bank number>
<Delimiter>
Acquire the current bank number.
BS
BS <Bank number>
<Delimiter>
OK <Delimiter>
Replace the current bank number by the
specified bank number.
ZR
ZR <Task number>
<Delimiter>
OK <Delimiter>
Execute a zero reset for the specified task.
If the <task number> is omitted, the zero
reset is executed for the task of which the
result is currently displayed.
If "4" is set for the <task number>, this is
executed for all tasks.
ZC
ZC <Task number>
<Delimiter>
OK <Delimiter>
Cancel the zero reset of the specified task.
If the <task number> is omitted, the zero
reset is executed for the task of which the
result is currently displayed.
If "4" is set for the <task number>, this is
executed for all tasks.
TM
TM <0:OFF/ 1:ON>
0<Delimiter>
OK <Delimiter>
Executes TIMING input.
* Calculates OR with the parallel input.
RT
RT <0:OFF/ 1:ON>
0<Delimiter>
OK <Delimiter>
Executes RESET input.
* Calculates OR with the parallel input.
LD
LD <0: Lit/ 1: Out>
0<Delimiter>
OK <Delimiter>
Turns the logical beam ON/OFF.
VR
VR <Delimiter>
<Model/Version>
<Delimiter>
Acquire the system version information.
(Example) ZW-C10 1.100 <Delimiter>
CA
CA 0 <Delimiter>
OK <Delimiter>
Calibrate the Sensor Head.
LS
LS <Save intervals>
<Number of saves> <Delimiter>
OK <Delimiter>
Start the internal logging of the data.
LE
LE <Delimiter>
OK <Delimiter>
End the internal logging of the data.
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......