![background image](http://html.mh-extra.com/html/omron/zw-c1-t/zw-c1-t_user-manual_1633390099.webp)
Setting the Output Conditions
ZW User's Manual
97
4
Settings for
Func
tion
Setting Triggers
Set how measurement start to end timing is to be input.
Item
Setting item
Setting value
Description
Trigger
Set-
tings
Trigger method
External input
Enter the trigger for the start of sampling in "TIMING input" for the 20-pole
terminal block.
The period that the signal input to "TIMING input" is ON is the sampling
period.
When a delay time is set, input OFF and sampling period end are
not synchronized. End timing is after the specified sampling period
has elapsed.
Selfup trigger
The period in which the measured value exceeds the preset self-trigger level
is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
When a delay time is set, the timing that the measured value falls
below the self-trigger level and the sampling period end are not
synchronized. End timing is after the sampling period has elapsed.
Selfdown trigger
The period in which the measured value falls below the preset self-trigger
level is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
When a delay time is set, the timing that the measured value
exceeds the self-trigger level and the sampling period end are not
synchronized. End timing is after the sampling period has elapsed.
Trigger level
-999.999999 to 999.999999
[mm]
(default value: 0)
Set the self-trigger level when [SELF-UP] or [SELF-DOWN] are selected at
[Trigger method].
Trigger
hysteresis
0 to 999.999999 [mm]
(default value: 0.05% of
measuring range)
Set the hysteresis width for the self-trigger when [SELF-UP] or [SELF-
DOWN] are selected at [Trigger method].
ON
OFF
Timing input
Sampling period
Important
Self-trigger
level
Measured
value
Sampling period
Hysteresis width
(for self-trigger)
Action point
Return point
Sampling period
Important
Self-trigger
level
Measured
value
Hysteresis width
(for self-trigger)
Action point
Return point
Sampling period
Important
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......