![background image](http://html.mh-extra.com/html/omron/zw-c1-t/zw-c1-t_user-manual_1633390260.webp)
Setting Threshold Value
258
ZW User's Manual
Operating procedure
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN
mode.
2
Press the
/
key to select MEAS, then press the
key.
3
Press the
/
key to select TASK1, then press the
key.
4
Press the
/
key to select JUDGE, then press the
key.
5
Press the
/
key to select TEACH, then press the
key.
6
Select HIGH THRESHHOLD as the threshold value type.
Press the
/
key to select H.JUDGE, then press the
key.
7
The current measured value is displayed on the main display and the
current set threshold value is displayed on the sub-display. (At this
time, the THRESHOLD-H indicator lights up.)
Press the
key to display the current measured value on the
lower line in the main display.
8
Press the
key to start a teaching.
The indicator for the current measured value on the main display
lights up, applying the value as a threshold value.
9
Press and hold the
key for two seconds to enter the RUN
mode.
Summary of Contents for ZW-C1*T