Displaying measured values in graphs
118
ZW User's Manual
Specifying the sampling start and end conditions
You can specify the conditions for starting and ending sampling.
1
Check the Trigger start conditions/Trigger end conditions checkbox.
2
Select the trigger condition.
3
Select the trigger target.
Item
Setting item
Range
Description
Sampling setting
Trigger start
conditions
Input and output
Specify parallel I/O (TIMING, ZERO, BUSY, ENABLE,
HIGH, PASS, LOW) as the trigger condition.
Data slope
Sets the change in the measured values for TASK1-4 as
the start condition. Sets the trigger level. When the
measured value rises above this value or falls below it, the
start trigger is issued.
Data window (In)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is inside the
range, the start trigger is issued.
Data window (Out)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is outside
the range, the start trigger is issued.
Trigger end
conditions
Input and output
Specify parallel I/O (TIMING, ZERO, BUSY, ENABLE,
HIGH, PASS, LOW) as the trigger condition.
Data slope
Sets the change in the measured values for TASK1-4 as
the start condition. Sets the trigger level. When the
measured value rises above this value or falls below it, the
end trigger is issued.
Data window (In)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is in the
range, the end trigger is issued.
Data window (Out)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is outside
the range, the end trigger is issued.
Number of data
The sampling data is counted from when the start
condition is met and when the specified number of data
points have been sampled, the end trigger is issued.
Item
Setting item
Range
Description
Trigger target
I/O
TIMING
TIMING input signal (parallel I/O)
ZERO
ZERO input signal (parallel I/O)
BUSY
BUSY output signal (parallel I/O)
ENABLE
ENABLE output signal (parallel I/O)
HIGH
HIGH output signal (parallel I/O)
PASS
PASS output signal (parallel I/O)
LOW
LOW output signal (parallel I/O)
Data slope
Data window (in)
Data window (out)
TASK1 TASK1
measurement
results
TASK2
TASK2 measurement results
TASK3
TASK3 measurement results
TASK4
TASK4 measurement results
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......