Recovering calibration ROM data
ZW User's Manual
129
5
Conve
n
ien
t Fu
nctions
5-6 Recovering calibration ROM data
If an abnormality occurs in the sensor's calibration ROM, you can recover the backed up calibration ROM data
into the sensor.
Explorer pane
: [Device Group] | [(Sensor Name)] (double click)
Edit pane
: [Tool] icon (
)
The calibration ROM data is different for each sensor serial number. Select the backup file that matches the sensor
serial number. Measurement will not be correct unless they match.
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.57
2
Click the [ROM recovery] in [Sensor setup].
Select the calibration ROM backup data file.
Important