Setting Sensing
ZW User's Manual
71
4
Settings for
Func
tion
Setting the Material of the Target to Measure
Setting the material of the target to measure.
Explorer pane
: [Bank] | [(Bank data name)] (double-click)
Edit pane
: [Sensing setting] icon (
)
Sensing settings window : [Operating mode]
Refer to p.286 for the linearity data of various materials according to each material setting.
Item
Setting item
Setting value
Description
Operation mode
Material
Normal (default value)
A measurement can be performed at specific linearity
regardless of the type of target to measure.
Usually, select this setting.
Mirror
Select this mode when specular reflection (regular
reflection) occurs on the surface. (Glass, wafer, lustrous
metal, etc.)
Rough
Select this mode when diffuse reflection occurs on the
surface.
1
Select the material of the target to measure
at [Material].
The material of the target to measure can also be set by the operating keys on the Sensor Controller.
Setting the Material of the Target to Measure p.228
Important
Note