Installation
ZW User's Manual
2
Install
a
tion an
d
C
onn
ecti
ons
35
• Handling fiber cables
Use them in compliance with the following.
-Fiber cable bend radiuses must be at least 20 mm.
-Do not let bending cause stress at the connecting section of a fiber connector.
-Do not yank hard on a fiber cable.
-Do not step on a fiber cable or place anything heavy on it.
• Do not touch the end surface of a fiber cable, or the cable may be degraded in performance. Should the end surface
be touched or soiled, wipe the dirt away using a commercially available cleaner exclusive for fiber or dry and soft
cloth. Do not use a cloth moistened with alcohol, or the dirt may be reattached.
We recommend the following product as the fiber cleaner.
• Use the optional ZW-XCL when cleaning the groove of the fiber connector.
• The fiber cable and fiber connectors should not be left with their protective caps removed, not even for a short
period of time. Leaving them unprotected can let dirt get on the end surface and cause performance deterioration.
• Calibrate the Sensor Head after removing and inserting a fiber cable.
Point
Item
Model
Manufacturer
Tip of fiber cable
OPTIPOP R1
ATC-RE-01
NTT Advanced Technology Corporation
Important
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......