Setting Measurement Items
ZW User's Manual
81
4
Settings for
Func
tion
Measuring the Thickness
If you have glass of known thickness, scaling can be adjusted referenced to that glass so that transparent
objects can be measured more easily.
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Measurement mode] - [Thickness]
Item
Setting item
Setting value
Description
Thickness
settings
Measurement
area
Area 1 (default value)/Area 2
The area targeted for measurement can be switched.
Measurement
surface 1
Edge1/Edge2/Edge3/Edge4/
peak (default value)
The measurement surface (top surface) targeted for
measurement can be switched.
The peak is the surface of all measurement surfaces having
the highest received light amount.
Measurement
surface 2
Edge1/Edge2/Edge3/Edge4/
peak (default value)
The measurement surface (rear surface) targeted for
measurement can be switched.
The peak is the surface of all measurement surfaces having
the highest received light amount.
1
Select the measurement area targeted for
measurement from [Measurement area].
2
Select the measurement surface (top
surface) targeted for measurement from
[Measurement surface 1].
3
Select the measurement surface (rear
surface) targeted for measurement from
[Measurement surface 2].
• Thickness measurement can also be set by the operating keys on the Sensor Controller.
• For the operation of the thickness scaling, refer to p.88.
Sensor head
Transparent
object
Edge 1st
Edge 2nd
Thickness
Note
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......