Connecting Parallel I/O
268
ZW User's Manual
Settings for Binary Output
Output Content Setting
As an example, here is an explanation of the procedure for outputting the measured value of TASK1 as binary
for each measuring cycle.
Operating procedure
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN mode.
2
Press the
/
keys to select "I/O" and press the
key.
3
Press the
/
keys to select "BINARY" and press the
key.
4
Press the
/
keys to select "OUTPUT" and press the
key.
5
Select the binary output contents.
The current set value is displayed on the sub-display.
Press
/
key to enter editing mode and the sub-display blinks.
In this example, select "MEAS1" and press the
key.
6
Press the
/
keys to select "TASK1" and press the
key.
7
Enable TASK1 binary output.
The current set value is displayed on the sub-display.
Press
/
key to enter editing mode and the sub-display blinks.
Select "ON" and press the
key.
8
Press and hold the
key for two seconds to enter the RUN mode.