116
CONNECTION AND SETTINGS
:
Setting Parameters
5) Lens holding time (sec.): 0.0 to 3.0
Factory setting: 0.0
This parameter is for setting the time period for the blocking arm to remain at rest at the low-
est position in the range of 0.0 to 3.0 seconds.
By setting the proper holding time, the lens cup attached condition can be checked when
blocking the super water repellent lens.
In the case of the setting of 0.0 to 3.0 seconds:
The blocking arm returns to the original position after it has stopped for the set period of time.
In the case of the setting of 3.1 seconds:
The parameter is set to Stop.
When blocking is performed, the blocking arm keeps stopped at the lowest position.
When the OK button for “Complete blocking” displayed on the screen is pressed, the blocking arm
returns to the original position.
[Parameter items and setting contents: hole set-
tings]
1) Slot coordinate mode: Center/Edge
Factory setting: Center
Selects the reference points either from
center or edge when entering the coordi-
nates of slotted holes or the distance
between the hole and notch.
2) Drill bit diameter: 0.50 to 5.00
Factory setting: 0.80
Set the parameter in accordance with the diameter of the drill of the lens edger.
The default value range is 0.50 to 5.00 mm in increments of 0.01 mm.
Center