125
CONNECTION AND SETTINGS
:
Connection
4
4.2.3
Setting connected instruments
For the instrument connected to the Ice 900, set the Communication Interface parameter according to
the system configuration as follows.
{
Setting the Lex 1000
*1: Never change the factory setting of Mac Address that had been set uniquely for each individual
Ice 900 before shipment. Do not change the factory settings.
*2: Set these parameters in accordance with the network to which the Ice 900 is connected.
*3: Set the parameters on the screen for servicing. Ask NIDEK or your authorized distributor for reset-
ting of the above parameters.
*: When the Ice 900 setting is Blocker VCA, contact NIDEK or your authorized distributor.
The Ice 900 has not displayed on the example
image of Lex 1000 Communication Setting
screen yet. However, use the Communication
Setting screen for settings.
Parameter
Blocker Lex
(when using RS-232C
port)
Mini Lab
(when using RS-232C
port)
Mini Lab
Blocker Lab
(when using LAN port)
Communication setting
(system)
Blocker Lex
(RS-232C)
Mini Lab
(RS-232C)
Mini Lab
(Ethernet)
Server ID No.
Ice 900
ID No.: 32
Ice 900
ID No.: 32
ID No. of the host PC (Ice
900): 32
My ID No.
ID No. of the Lex 1000:
(Any number except for the
Server ID No.)
ID No. of the Lex 1000:
(Any number except for the
Server ID No.)
ID No. of the Lex 1000:
(Any number except for the
Server ID No.)
Mac Address *3
–
–
Never change the factory
setting. *1
IP address
–
–
Set the parameter uniquely
for each individual ME-1000
*2
Subnetmask *3
–
–
Set the parameter uniquely
for each individual ME-1000
*2
Default Gateway *3
–
–
Set the parameter uniquely
for each individual ME-1000
*2
Host IP address
–
–
IP address of the host PC
(Ice 900)
RS-232C Baud Rate (COM1) *3
Set the Baud Rate to the
same value as the Ice 900
Set the Baud Rate to the
same value as the Ice 900
–
Blocker Lex
Host ID No.
(ID No. of the Ice 900)
ID No.