127
CONNECTION AND SETTINGS
:
Connection
4
{
Setting the SE-9090 Express
*1: Never change the factory setting of Mac Address that had been set uniquely for each individual
Ice 900 before shipment.Do not change the factory settings.
*2: Set these parameters in accordance with the network to which the Ice 900 is connected.
*3: When the Ice 900 or NIDEK-produced PC software is used, set the host port No. to “55555”. When
the other software is used, set the host port No. to the port No. of the used software.
*: When the Ice 900 setting is Blocker VCA, contact NIDEK or your authorized distributor.
*: Blocker Lex cannot be connected.
Parameter
Mini Lab
(when using the RS232C port)
Blocker Lab
Mini Lab
(when using the LAN port)
Communication interface
LAN
LAN
COM port/Protocol
RS232C
TCP
Baud rate
Set the Baud Rate to the same value as
the Ice 900.
–
Mac address
–
Never change the factory setting. *1
IP address
–
Set the parameter uniquely for each
individual SE-9090 Express. *2
Subnet mask
–
Set the parameter uniquely for each
individual SE-9090 Express. *2
Gateway
–
Set the parameter uniquely for each
individual SE-9090 Express. *2
Host IP address
–
Server (or Ice 900) IP address
Host ID
Ice 900
ID No.: 32
Server (or Ice 900)
ID No.:32
My ID
SE-9090 Express ID No.
(Any number except for the Server No.)
SE-9090 Express ID No.
(Any number except for the Server No.)
Port No.
–
55555 *3