89
OPERATING PROCEDURES
:
Drilling Settings
2
7
When the measurement is completed,
the detected outline is displayed in
white solid line and hole is displayed in
pink dashed line on the screen.
When the outline or marking cannot
be detected, press the
button to
perform the operation from step 6.
8
When the measurement is completed,
press the DBL button to enter a
numeric value.
9
Press the
button to return to the Hole Editor screen.
Pressing the
button confirms the shape loaded result.
10
Add the hole that cannot be measured.
Or delete the hole measured by mis-
take.
The hole is displayed in red in the
specified position.
Rough positioning is acceptable
because it will be precisely aligned
later.
When dragging is performed with the
screen pressed, the hole moves with
the dragging.
Press the hole delete
button to
delete the circle displayed in red.
Cancel
Cancel
• If “Measurement completed. Failed to align axis.” is displayed on the information bar, axis
alignment failed. Check that the WD setting of the single vision lens is fit in the distance
between markings on the right and left of the Layout screen, perform measurement again.
When the same error occurs if WD is set, manually align the angle properly with the manual
alignment mode.
For details, see “2.7.6 Measuring demo lenses (Manual Alignment mode)” (page 94).
• After completing the measurement, the screen cannot be changed to the screen other than
the Hole Editor screen. After pressing the Cancel button and canceling the measurement
result, changing to other screen becomes possible.
OK
OK
Hole add buttons
Red circle