128
CONNECTION AND SETTINGS
:
Connection
{
Setting the ME-1000
*1: Set the parameter in the Communication Setting screen.
*2: Set the parameter in the Network Setting screen.
*3: Never change the factory setting of Mac Address that had been set uniquely for each individual Ice
900 before shipment. Do not change the factory settings.
*4: Set these parameters in accordance with the network to which the Ice 900 is connected.
*: When the Ice 900 setting is Blocker VCA, contact NIDEK or your authorized distributor.
*: Blocker Lex cannot be connected.
{
Setting of the LT-900
To directly connect the LT-900 to the Ice 900, connect the LT-900 to the TRACER port of the Ice 900.
*1: Changing the settings needs a PC and Tracer Utility software.
*2: If the SW1 RATE switch on the side circuit board (BA15) of the tracer is upward, Baud Rate is
9600, downward, BAUD RATE is 38400.
Parameter
Mini Lab
(when using the RS232C
port)
Mini Lab
(when using the LAN
port)
Blocker Lab
*1
Communication port
RS-232C
Ethernet
Ethernet
*1
Communication interface:
LAN
LAN
LAN
*1
Baud rate
Set the Baud rate to the same
value as the Ice 900
–
–
*1
Protocol
–
TCP
TCP
*1
Host ID
Ice 900
ID No.: 32
Ice 900
ID No.: 32
ID No. of the host PC: 32
ID No.: 32
*1
My ID
ID No. of the ME -1000:
(Any number except for the
Server No.)
ID No. of the ME -1000:
(Any number except for the
Server No.)
ID No. of the ME -1000:
(Any number except for the
Server No.)
–
Mac address
–
Never change the factory
setting. *3
Never change the factory
setting. *3
*2
IP address
–
Set the parameter uniquely
for each individual ME-1000
*4
Set the parameter uniquely
for each individual ME-1000
*4
*2
Subnet mask
–
Set the parameter uniquely
for each individual ME-1000
*4
Set the parameter uniquely
for each individual ME-1000
*4
*2
Gateway
–
Set the parameter uniquely
for each individual ME-1000
*4
Set the parameter uniquely
for each individual ME-1000
*4
*2
Host IP address
–
IP address of the Ice 900
ID No. of the host PC
IP address
Parameter
Mini Lab
*1
Protocol
STD
*2
Baud Rate
Set the Baud rate to the same value
as the Ice 900.