65
OPERATING PROCEDURES
:
Setting Processing Conditions and Entering Layout Data
2
5
Enter the size (enlarged or reduced lat-
eral based size of lens shape)
.
Enter an arbitrary value in the Size field
if necessary.
Press the Size numeric button to dis-
play the ten-key pad. Enter any value
and press the
button.
The size value range is -9.95 to +9.95 mm in increments of 0.01 mm. The minus mark indicates
that the lens shape is reduced and the plus mark indicates that the lens shape is enlarged.
The numeric value can also be entered by the jog dial.
The numeric value can be entered in increments of 0.05 mm.
For details, see “1.3.8 Ten-key pad screen” (page 25).
The lens shape is enlarged or reduced laterally
while maintaining a similar form. At this time, the
optical center is regarded as the reference posi-
tion.
For example to the right: Size = +5 mm:
6
Select Active/Passive (Layout mode).
• The default displayed on the optical center height numeric button can be specified by
setting the Optical center height parameter.
Enlarged shape size
Original shape size
Original shape size +5 mm
• This function allows one pattern to be used for several different-sized patterns.
• Changing the shape to a different form is also possible.
For details, see “2.6 Shape Editor Function” (page 75).