maIntenance
113
Check following values:
Front and Back Gain
≤ 8
Ref Gain
≤ 16
4. If the gains are over the limit, refer to chapter 9 “Troubleshooting”.
8.4.2 WASHER VOLUME CALIBRATION
The washer volume should be between 500 and 700 µl. The calibration will take
approximately 3 minutes. The screen will show the pump steps settings and
new pump steps required for system delivery in all four wash steps.
1. Perform a System Flush (refer to 8.2.4).
2. Open the Maintenance scheduler status screen and press
Perform
behind
the Washer Calibration function.
3. The Calibration will start automatically and a pop up window will appear
when it finished.
4. Press
Ok
to get the screen, displayed below:
FIGURE 109
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......