maIntenance
105
8 MAINTENANCE
A regular maintenance is necessary, to keep the system in good operating con-
dition. The scheduler will help to manage the daily, weekly and monthly main-
tenance. From time to time it’s also necessary to do some maintenance on de-
mand i.e. Cuvette change.
A recommended summary of tasks for daily, weekly and monthly maintenance
is presented in the table below:
Frequency
Task
Daily
Inspection and Cleaning of Probes
Check and Replace System Solutions
Empty Waste Bottle
Hydraulic Testing - System Flush
Weekly
Intensive Cuvette Cleaning
Cuvette Water Blank
Service Backup
Monthly
Photometer Calibration
Washer Volume Calibration
Clean System Bottles
Intensive Washer Cleaning
On Demand
Cuvette Change
Lamp Replacement
Pump Tube Replacement
8.1 Scheduler
Maintenance Scheduler can be accessed either from the warnings bar (if a warn-
ing is yellow or red, click on it) on the system information field at the main win-
dow or from main menu:
FIGURE 103