HumaStar 600
| User manual
112
1. Open the Maintenance scheduler status screen and press
Perform
behind
the Service Backup function.
2. The Backup starts automatically and a pop up window appears when it fin-
ished.
3. Press
Ok
and the scheduler will be updated.
8.4 Monthly Maintenance
Proceed to the weekly maintenance routine.
8.4.1 PHOTOMETER CALIBRATION
This calibration will determine the optimum reading position in the middle of
each cuvette.
Please perform this maintenance directly after intensive cuvette clean and cu-
vette water blank to ensure the cuvettes in Pos 1 are clean. The calibration will
take approximately 10 minutes.
1. Open the Maintenance scheduler status screen and press
Perform
behind
the Photometer Calibration function and a pop up window will appear. If the
cuvette is clean, a new placement is not necessary.
2. Press
Ok
and the calibration will start automatically.
3. When finished refer to the summarize tab and following window will be dis-
played: