maIntenance
111
8.3.2 CUVETTE WATER BLANK
Cuvette water blank should be done weekly and after each intensive cuvette
cleaning.
Washers are dispensing system wash in all cuvettes and photometer records
the background measurement of each cuvette at all 12 filters. The process takes
places into two parts; at first system wash is dispensed to the first 40 cuvettes,
absorbances are read, cuvettes are then dried and then the other half of the
cuvette tray is filled and read. Process takes approximately 15 min and if accept-
ance criteria are fulfilled, cuvettes are considered clean.
1. Open the Maintenance scheduler status screen and press
Perform
behind
the Cuvette Water Blank function.
2. The cuvette water blank starts automatically.
3. After the cuvette water blank finished the scheduler will be updated auto-
matically. The actual and the previous results are displayed on the screen.
8.3.3 SERVICE BACKUP
Perform the backup at least once per week to save all necessary information
needed to reinstall the system and retrieve data in an emergency.
Save the Backup on an external memory device. Duration can vary from a few
seconds up to a couple of minutes, depending on the size of the database.
FIGURE 108