5.4.3 ORDERING A CALIBRATION
To load the calibrator‘s cups on the instrument sample tray refer to section 5.2.6.
5.4.4 CALIBRATION ACCEPTANCE
Open the
Calibration
Menu and select on the left side the
Pending Accept-
ance
Submenu. Following window will be displayed:
To confirm a calibration test result, select the desired method and the calibra-
tion curve including replicates will be displayed.
Before acceptance of a calibration indivdual replicates can be disabled. Unselect
the box in front of a replicate which should be excluded from the calibration
curve calculation.
Column to the right of each function will show the least squares adjusting val-
ue. Select function with minimum value unless some special feature is required.
Experiment deselecting one or more standards and observe recalculated values.
Press
Accept
once the calibration curve and values are as desired. In this case,
at least one value for each vial must be selected.
Press
Reject
to mark the calibration as unusable.
5.4.4.1 Flagged results
Press
Detail
to expand the detailed area for the selected reaction. This
panel points out active flags. Results may be flagged if validity limits, duplica-