6.2.3 QUANTITATIVE
Calibration, reagent blank and reaction specific parameters are set up in the
Quantitative tab.
Item
Input
Comments
Method Correlation
The concentration value (X) will be corrected
with the equation Y=AX + B
Slope
-999999 to 999999
Slope = A
Offset
-999999 to 999999
Offset = B
Direction
Reaction direction according the method
Direction
Ascending or Decreasing
Select from drop down menü
Direction Check
Check / No Check
Direction will be checked and result will be
flagged if direction is wrong and the result
has to be manually accepted
Calibration type
Curve / Linear
Check / No Check
Check if one or more calibrators should be
used
Validity time (days) Check / No Check;
1 to 999999
Check and insert the number of days, if the
calibration stability should be checked; after
the number of days or if a new reagent lot is
on board, the calibration curve expires and a
new calibration will be requested.
Fixed
Check / No Check; 1 to 20
Check and insert the number of points used
in the calibration; If this option is not select-
ed, the user can define a calibrator set with
any number of points.
FIGURE 87
TABLE 12