HumaStar 600
| User manual
90
Item
Input
Comments
Concentration
validity limit
Linearity range check
Low
-999999 to 999999
Insert the low linearity limit according the
method IFU
Not rerun, Rerun,
More Sample
Select from drop down menu; result can be
flagged, repeated as original run or repeated
with more sample volume
High
-999999 to 999999
Insert the high linearity limit according the
method IFU
Not rerun, Predilution,
Less Sample
Select from drop down menu; results can
be flagged, repeated with a predilution (the
predilution factor is established automat-
ically (taking in account low and high lim-
it and calculates the dilution factor for the
measured absorbance to get readings in
the middle) or repeated with less sample
volume
Integrity check
absorbance limit
Reagent integrity (quality) check; only reagent
will be dispensed into a cuvette and measured
each morning
Low limit
-999999 to 999999
Insert a low absorbance limit
High limit
-999999 to 999999
Insert a low absorbance limit
Concentration
duplication limit
Result verification check; normal repetion can
be triggered according to each laboratory need
Low limit
-999999 to 999999
Insert a low absorbance limit
High limit
-999999 to 999999
Insert a low absorbance limit
Reagent blank
automatic
acceptance
If automatic acceptance is enabled, reagent
blank results within the specified limits will
not appear on the Pending Acceptance window
of blanks, they show up directly in the In Use
window of blanks
Low limit
-999999 to 999999
Insert a low absorbance limit
High limit
-999999 to 999999
Insert a high absorbance limit
Calibration
factor automatic
acceptance
If automatic acceptance is enabled, calibration
factors within the specified limits will not ap-
pear on the Pending Acceptance window of
calibration, they show up directly in the In Use
window of calibrations
Low limit
-999999 to 999999
Insert a low factor limit
High limit
-999999 to 999999
Insert a high factor limit
Change in %
Check / No Check;
-999999 to 999999
Check if the deviation from the previous valid
(in use) calibration should be checked and in-
sert the allowable percentage
TABLE 13
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......