3 SYSTEM DESCRIPTION
3.1 Unpacking
Remove all the parts from their package.
When unpacking the instrument, please make sure that the following items are
contained in the packing. In case of damage or missing item, please contact the
supplier immediately.
Quantity
Description
[REF]
1
Software CD
16661
2
Reaction cuvettes (box of 1200)
16661/1
2
Drying block kit
16661/11
1
Reagent recipients with cap (x 30) vol. 70 ml
16661/2
1
Reagent recipients with cap (x 30) vol. 25 ml
16661/3
2
Peristaltic Pump tubing kit x 3
16661/4
2
Sample tubes 13 mm. kit x 100
16661/5
1
Halogen lamp 12V 20W
16661/7
5
Sample Rack
16661/15
1
Serial Cable
16661/16
1
User Manual
16660/001
3.2 Installation
3.2.1 INSTALLATION REQUIREMENTS
Carefully read the safety instructions included in this manual.
Install the instrument on a hard floor with a resistance of at least 50 kg/cm
2
; use,
if possible, ceramic or stone floor.
Avoid carpets or very soft rubber.
Mains should be close to the instrument (less than two meters) and must fulfill
local regulations.
Free access to main switch is required. A distance of 50 cm from the left side
of instrument to nearest table or wall is advisable. Right side must have a free
space of at least 30 cm for ventilation purposes.
Space must be empty over instrument to 2.10 m. Avoid using shelves, walls or
screens above instrument.
Instrument is mounted on wheels and can be moved towards the front for ser-
vicing and cleaning purposes. Allow free space of about two times the instru-
ment depth.
TABLE 1
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......