defInItIon of metHodS
89
Mandatory
formula
None, Linear, MultLinear,
Sigmoidea, Logit4, Logit5 or
Spline.
Select from dop down menu if a special for-
mula should be used; if none is used the for-
mula can be manually modified in the Cali-
bration>Pending Acceptance window
Factor
-999999 to 999999
Insert the factor
Use from this
method
Check / No Check
Check if the calibration curve from different
method should be used and select the meth-
od from the drop down menu
Reagent blank
Enabled
Check / No Check
Check if a reagent blank should be used in
the calculation of the result
Sample
replacement with
Diluent/Reagent
Select which solution should be used
Diluent
Drop down menu
All diluents defined in Method Definition >
Solutions are available
Reagent
Drop down menu
All reagents defined in the Main tab are
available
Validty time
1,3,6 or 12 hours;
1 to 120 days
New reagent blank will be requested if the
defined time expires or a reagent lot or bot-
tle number is modified
Default replicates
1 to 10
Automatically
performed
Check / No Check
Check if the reagent blank should be per-
formed automatically
6.2.4 LIMITS
Checks and automatic acceptance ranges are set up in the Limits tab.
FIGURE 88