defInItIon of metHodS
93
Tip pre wash
Check / No Check
Check to enable an extra probe wash before
aspirating the reagent
Water
Check / No Check
Check to use water for the extra wash
Solution
Check / No Check,
drop down menu
Check and select the solution which should
used for the extra wash
Reagent
Check / No Check,
drop down menu
Check and select the reagent which should used
for the extra wash
When
Always / After interf.
methods
Usage always or only when a defined interfer-
ing mehtod was used before the method
Cuvette post wash
Check / No Check
Check to enable the addition of a solution to
the reaction cuvette before washing
Solution
Drop down menu
Select the solution to be used
Vol (µl)
0 to 50
Insert the volume to be used
Time (sec)
0 to 60
Insert the time the added solution should be
incubated together with the reaction liquid
After sample tip wash
Check / No Check
Check to enable an external tip wash after aspi-
ration of the sample (to improve the lineartiy in
methods with very low absorbances)
Interfering methods
Add or delete methods which interfer the
existing method
Sample Acceptance
Manual/Automatic
Manual to send all results to the Pending Ac-
ceptance tab in the Tests Menu; automatic to
send all results without flag to the Done tab in
the Tests Menu (flagged results will all the time
be show up in the Pending Acceptance)
Arm selection
Select the arm which should be used (important
to take contamination avoidance and speed in
consideration)
Front
Check / No Check
Check if the front arm should be used
Back
Check / No Check
Check if the back arm should be used
Shake
Off / Normal / x2 / x3/
x4
Adjust the shaking during the tip cleaning, in-
termediate wash, reaction cuvette and pre dilu-
tion mixing. Normal (centrifugal) mixing dura-
tion is 500 msec
Viscouse reagent
Check / No Check
Check if the reagent has a sticky cosistensy and
the dilutor speed will be slowed down
On board stability
Time
1,3,6 or 12 hours;
1 to 120 days
Select the on board stability given by the man-
ufacturer
Keep using
Check / No Check
Check to enable the reagent longer then the on
board stability (a warning will be issued)
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......