72
erated by using the Control Set Name. This Id can be modified at the moment of
generation but not at a later time.
New tests included in the Control Set can be added later using the Control sub-
menu: Select the control from the list on the right and press
Add Test
. Check
one or more tests that you want to add. Press
Ok
when done or
Cance
l
to
abort.
To remove a test from the control press
Delete Test
. Select one or more tests
from the list. Press
Ok
to confirm or
Cancel
to abort.
5.6.3 PROCESSING A CONTROL
Once the control has been requested, load it in a vial and place in the instrument
as indicated in 5.2.6.
If the barcode reading procedure is performed and a control set is recognized, it
will be associated with any pending action already defined in the scheduler. If
not programmed in scheduler, no action will be taken on the control.
Use the
Pending
Submenu tests to review non-processed (idle) or in-progress
control tests.
User must consider that test results may stay idle if the system operation runs
out of reagent or control sample.
Use the
Cumulative Historic
Submenu to review data already stored in previous
runs. They will be organized in directories, where results of each month are au-
tomatically stored.
5.6.4 PROCESSED CONTROLS
To display processed Controls, open the
Quality Control
Menu and select on the
left side the Submenu
Done
.
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......