The option gives access to review in real time any reaction. This method does not
run with others as part of the automatic cycle. It is used alone, sample by sample
and its purpose is to study slope, end point, optimum range, incubation period,
etc. of individual reactions.
In
Options
, basic method parameters are defined: wavelengths, number of
reagents, volumes, total measuring time. In Calibration page, analysis type is
defined and fit formula calculated. In Results, measurements can be viewed
graphically and Time 1 and Time 2 can be adjusted in order to focus on required
time interval. This funtion is mostly used for research and development projects
as well as method optimization.