HumaStar 600
| User manual
58
5.4 Calibration
To enter a new calibration or define a calibrator set, press
Calibration
Menu button.
In the Calibration Menu following options are available:
Calibration
New calibrations can be requested or requested calibrations
can be deleted.
Pending
All non-processed and in-process calibrations are shown.
Pending
Acceptance
In case the manual acceptance is programmed and/or the reac-
tion is flagged the calibration has to be manually accepted. Use
the reject function to rerun the calibration and Detail to see the
reaction measurements.
In Use
Actual calibrations can be reviewed, deleted or printed
Historic
Former calibrations can be reviewed or reactivated for the ac-
tual use. The calibration is shown in the In Use screen with a
warning and the results should be flagged (function “Check Re-
used Calibration” in
Maintenance>Parameters>Software>Gen-
eral
has to be enabled). WARNING this funtion must be cau-
tiously used, never mix calibrations and actual reagent lots.
ISE
Actual and historic ISE calibrations can be reviewed.
Automatic
calibration
Calibrator Lot numbers can be defined for the use in the next
automatic calibration, this can be in advance.
Calibrator Set
Single and Multipoint calibrators can be set up, modified or
deleted.
FIGURE 57
TABLE 8
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......