routIne taSKS
73
In the upper section following filter options are available to display the desired
data:
Date (From - To)
Index (From - To)
Test Id
Lot Number
Control Id
Last (number of displayed results)
Date selectors show calendar for simple selection. All results are indexed and in-
dex is shown at the beginning of each column. Selectors can be used separately
or together. Selectors act as logical AND which allows refining the selection. The
result are dislayed including the low and high limits and they are marked in the
last column by an X when they are out of range.
Select a result and press the
Details >>
button to show extended information.
The raw data curve is displayed including the information about dilution factor,
reagent- and calibrator lot numbers. Press the
<< Plot
button to display the re-
sults in a graph.
FIGURE 72