tion limits or reference class limits and so on are exceeded. Both low and high
limits can be set independently (see chapter 6.2.4 for details).
5.4.4.2 Calculations
For the calibration curve calculation following formulas are available:
Linear
Abs = Ax + B
Multilinear
Cx = C1s+(C2s-C1s)*(Sx-S1s)/(S2s-S1s)
linear interpolation between consecutive standards
Spline
C = a*A3 + b*A2 + c*A + d
where consecutive 3d degree polynomials joining consecutive
data
Sigmoid
Abs = L + ((H – L)/(1 + exp(-(Conc – a)/b)))
where H, L , a and b are automatic adjusting parameters
Logit4
Abs = R + (K/(1 + exp(-(a+b*ln(Conc)))))
where K, a,b are automatic adjusting parameters
Logit5
Abs = R + (K/(1 + exp(-(a+b*ln(Conc) + c* Conc))))
where K, a,b,c are automatic adjusting parameters
Calibration fit parameter is shown in all cases.
If curve is forced to pass through zero, functions logit 4 and logit 5 will diverge
because of the logarithm function and will not be shown.
Logit5 requires a minimum of 5 standards; logit4 requires a minimum of 4 stand-
ards; sigmoid and multi linear functions require a minimum of 3 standards.
5.4.5 AUTOMATIC CALIBRATION
It is possible to define for each method, which Calibrator Set will be used in
the next calibration, once the present one is expired, deleted or a lot number
changed. This set can also be “Run in advance” and be ready several hours be-
fore the present calibration expires. Open the
Calibration
Menu
and select on
the left side the
Automatic Calibration
Submenu.
TABLE 9
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......