routIne taSKS
57
5.3.1 ACCEPTANCE OF RESULTS
To confirm the result for a given test open the
Pending Acceptance
Menu
, then select the desired sample test from the list and then press
Accept
. Press
Reject
to reject the test result. To reprocess the reaction just press
Rerun
or
Diluted Replicate
if additional dilution could be useful for confirmation of result.
A dilution factor can be introduced.
5.3.2 REFLEX TESTS
Reflex tests are those automatically launched when a given test is out of deter-
mined limits. Condition for the reflex test launch can be relative to fixed values
or to a reference class.
5.3.3 PRINTOUT OF RESULTS
Results can be printed in different ways and operator has full control on the
display of the results. This is a post run operation and is not related to the online
printing controlled by
Maintenance > Parameter > Software
.
Press the
button from Test Menu Done screen and following submenu
opens:
Printing can be performed on all samples, some filtered or selected. Se-
lection can be made by pressing
Ctrl
Key while pointer is pressed on selected
samples. A range is selected by pressing mouse on the first and then pressing
Shift
key and pointing to the last one of the desired ranges.
There are three types of reports: Continuous - samples are printed out one after
the other; 3 per page - according a fixed report format; Custom - according re-
port made by use of the Report Format Generator, refer to section 3.2.7.2 Modify
reports.
5.3.4 CUVETTE
Report use for service engineers only.
FIGURE 56
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......