Define control set id, introduce lot number and expiration date for traceability
purposes and filtering options.
To edit the definition of a already defined control set and select the control set
from the list on the right and then press
Edit
.
To add a new test to the control set press
Add Test
and select or type in the
method ID. Then type in the concentration range for that method and press
Ok
to add the test or
Cancel
to abort. To remove a method from the control
set, press
Delete Test
.
Define the number of default
Replicas
for each test within each control set.
This number will be used every time the control set is used or scheduled.
Repeat this operation for each required test on the profile. Then press
Ok
to
finish or
Cancel
to abort.
Alternatively, double click on the Quick load list to add tests in the control set
and type the values in the corresponding fields. Mean values are calculated
automatically.
Once tests are loaded in the grid, concentration limits can be edited at all times.
To remove a control set from the list, first select the control profile from the
list on the right and then press
Delete
. Press
Yes
to confirm or
No
to abort.
Delete All
button will remove all control sets.