routIne taSKS
59
5.4.1 CALIBRATOR SETS
Calibrators for single or mutipoint calibrations are defined for methods. This
calibrators can contain a single- or multiple methods. Following information are
included in the sets: calibrator set Id, Lot number, number of vials, tests, mean
values and automatic dilution if necessary. Once the calibrator sets are created
it is easy to request calibrations in routine. Only Supervisor and Power user can
define a new calibrator set.
5.4.1.1 Defining a calibrator set
Open the
Calibration
Menu and select on the left side the
Calibrator Set
Submenu.
To edit the definition of an already defined calibrator set, first select the
calibrator set from the list on the right and then press
Edit
.
To enter or define a new calibrator set, press
New
. Assign a calibrator Set Id and
the corresponding lot number of the calibrator. In case of a multipoint calibra-
tion insert the number of vials, default it is on 1 for a single point calibration. To
add a new test to the calibrator set press
Add Test
.
FIGURE 58