routIne taSKS
65
Select from the right window the desired Method-Calibration Set combination
and press the
Add
button. With the corresponding button the Run in ad-
vance interval is also fixed. This interval is checked every 15 minutes when the
system is running, it is in the “don’t stop” mode or during start operation.
5.5 Reagent Blank
Reagent Blank can be measured in the automatic procedure or
directly from the
Reagent Blank
Menu.
When the automatic measurement starts, blanks are requested for the follow-
ing reasons:
1. No reagent blank available
2. It expired
3. Reagent was removed, refilled or changed
4. A new reagent lot was introduced
In that case, a window opens with the list of required blanks. User can include
the number of required replicates.
FIGURE 64