HumaStar 600
| User manual
44
- Method name
- Reagent number
- Lot number
- Bottle number
- Expiration date
- Available test number
- Available volume
- Pending tests
- Date reagent was placed onboard
- Expired date
It can be defined morethan one vial for each method.
If the first bottle is empty, the system will automatically switch to the next
bootle. The + in the right field indicates the second bottle. Multiple lots of the
same reagent can also be placed simutaneously on the reagent tray and sepa-
rately calibrated.
The colors used for positions allow to easily distinguishing as follows:
Green
Reagent in position and in use (programmed samples)
Red
Reagent not calibrated (only shown if “Calibrated” on the “View”
is selected)
Blue
Tip cleaning solutions, diluents and C-Clean
Yellow
Reagent not in use
Light Gray
Free position
Dark Gray
Reagent not in use or removed
5.1.2 LOADING BARCODED REAGENTS, DILUENT AND CLEANING SOLUTIONS
Barcoded reagents, diluent and cleaning solutions can be randomly placed
into the reagent tray and will be checked and identified by pressing
BCR All
button. To load barcoded reagents on a single position right click a suitable posi-
tion and pick
Change & BCR check.
A question mark symbol will be issued at the
specified position. Place the reagent on this position and press
Apply Changes
to start the positioning sequence for reagent loading into the tray and barcode
reading. When a reagent is not included in the table of methods in use but its
code detected as located in the tray, it is automatically included in the methods
in use. If a reagent barcode label cannot be read a window will open and the
barcode can be insert manually.
TABLE 5
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......